Connect the power supply.2. Connect a precision resistance box (called a “6-dial” in this manual) to theplatinum resistance thermometer input terminals, as shown in the following diagram.9E5CNInput power supply10A 3B' 46-dialB 510E5CN-UInput power supply11A 3B' 26-dialB 11E5AN/ENInput power supply2A 16B' 176-dialB 18adj30Input type 0:p390e20cInput types 1, 2, 3, 4:p280e26bp-1045431903. Turn the power ON.4. Move to the calibration level.This starts the 30-minute aging timer. This timer provides an approximatetimer for aging. After 30 minutes have elapsed, the No. 2 display changesto 0.
You can advance to the next step in this procedure even if 0 is notdisplayed.5. Execute calibration for the main input.Press the M key to display the count value for each input type.The No. 2 display at this time shows the currently entered count value inhexadecimal. Set the 6-dial as follows:• Input type 0:390 Ω• Input type 1, 2, 3 or 4: 280 ΩAllow the count value on the No. 2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No.
2 display willflash and the count value will not be temporarily registered.6. When the M key is pressed, the status changes as shown to the left.Set the 6-dial to 10 Ω.Allow the count value on the No. 2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No. 2 display willflash and the count value will not be temporarily registered.Platinum Resistance Thermometer Calibration (Thermocouple/Resistance Thermometer Input)Section 6-47. Next calibrate the B-B' input.Change the connections as follows:9E5CNInput power supply10A 3B' 46-dialB 510E5CN-UInput power supply11A 3B' 26-dialB 11E5AN/ENInput power supply2A 16B' 176-dialB 18pb104544pb-04083strno8.
When the M key is pressed, the status changes as shown to the left.Set the 6-dial to 10 Ω.Allow the count value on the No. 2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No.
2 display willflash and the count value will not be temporarily registered.9. When the M key is pressed, the status changes as shown to the left.Short-circuit the 6-dial terminals to set 0 Ω.Note The 6-dial terminals must be short-circuited, because it is otherwiseimpossible to set 0 Ω for the 6-dial.Allow the count value on the No.
2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No. 2 display willflash and the count value will not be temporarily registered.10. When the M key is pressed, the status changes as shown to the left.The data to be temporarily registered is not displayed if it is not complete.Press the U key. The No. 2 display changes to yes. Release the key andwait two seconds or press the M key.
This stores the temporarily registered calibration data to EEPROM.To cancel the saving of temporarily registered calibration data to EEPROM, press the M key (while no is displayed in the No. 2 display) without pressing the U key.11. The calibration mode is quit by turning the power OFF.191Section 6-5Analog Input Calibration (Thermocouple/Resistance Thermometer Input)6-5Analog Input Calibration (Thermocouple/ResistanceThermometer Input)In this example, calibration is shown for a Controller with a Thermocouple/Resistance Thermometer Multi-input, with an analog input (0 to 50 mV) set asthe input type.E5CNE5CN-UE5AN/EN910Input power supply1Input power supply10Input power supply112STV4 −STV2 −STV17 −DMM5 +DMM1 +DMM18 +1,2,3...
1. Connect the power supply.2. Connect an STV and DMM to the analog input terminals (same as thermocouple inputs), as shown in the figure above.adj30t-54b9a5t--62988strno3. Turn the power ON.4. Move to the calibration level.This starts the 30-minute aging timer. This timer provides an approximatetimer for aging. After 30 minutes have elapsed, the No. 2 display changesto 0. You can advance to the next step in this procedure even if 0 is notdisplayed.5.
When the M key is pressed, the status changes as shown to the left.The No. 2 display at this time shows the currently entered count value inhexadecimal. Set the STV to 54 mV.Allow the count value on the No. 2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No. 2 display willflash and the count value will not be temporarily registered.6. When the M key is pressed, the status changes as shown to the left.Set the STV to −6 mV.Allow the count value on the No. 2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No.
2 display willflash and the count value will not be temporarily registered.7. When the M key is pressed, the status changes as shown to the left.The data to be temporarily registered is not displayed if it is not complete.Press the U key. The No. 2 display changes to yes. Release the key andwait two seconds or press the M key. This stores the temporarily registered calibration data to EEPROM.To cancel the saving of temporarily registered calibration data to EEPROM, press the M key (while no is displayed in the No. 2 display) without pressing the U key.8.
The calibration mode is ended by turning the power OFF.192Section 6-6Calibrating Analog Input (Analog Input)6-6Calibrating Analog Input (Analog Input)6-6-1Calibrating a Current InputIn this example, calibration is shown for a Controller with an Analog Input, witha current input set as the input type.1,2,3... 1. Connect the power supply.2. Connect an STV and DMM to the current input terminals, as shown in thefollowing diagram.E5CNE5AN/EN19Input power supplyInput power supply10adj30a-20c8e7a--14677strno2STV3 +STV16 +DMM4 −DMM17 −3.
Turn the power ON.4. Move to the calibration level.This starts the 30-minute aging timer. This timer provides an approximatetimer for aging. After 30 minutes have elapsed, the No. 2 display changesto 0. You can advance to the next step in this procedure even if 0 is notdisplayed.5. When the M key is pressed, the status changes as shown to the left.The No. 2 display at this time shows the currently entered count value inhexadecimal. Set the STV to 20 mA.Allow the count value on the No. 2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No.
2 display willflash and the count value will not be temporarily registered.6. When the M key is pressed, the status changes as shown to the left.Set the STV to 1 mA.Allow the count value on the No. 2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No. 2 display willflash and the count value will not be temporarily registered.7. When the M key is pressed, the status changes as shown to the left.The data to be temporarily registered is not displayed if it is not complete.Press the U key. The No.
2 display changes to yes. Release the key andwait two seconds or press the M key. This stores the temporarily registered calibration data to EEPROM.To cancel the saving of temporarily registered calibration data to EEPROM, press the M key (while no is displayed in the No. 2 display) without pressing the U key.8. The calibration mode is ended by turning the power OFF.193Section 6-6Calibrating Analog Input (Analog Input)6-6-2Calibrating a Voltage InputIn this example, calibration is shown for a Controller with an Analog Input, witha voltage input set as the input type.1,2,3... 1.
Connect the power supply.2. Connect an STV and DMM to the voltage input terminals, as shown in thefollowing diagram.E5CNE5AN/EN91Input power supplyInput power supply10adj30Input type 2 or 3:1v-5c7c3STV4 −STV17 −DMM5 +DMM18 +3. Turn the power ON.4. Move to the calibration level.This starts the 30-minute aging timer. This timer provides an approximatetimer for aging.
After 30 minutes have elapsed, the No. 2 display changesto 0. You can advance to the next step in this procedure even if 0 is notdisplayed.5. When the M key is pressed, the status changes as shown to the left.The No. 2 display at this time shows the currently entered count value inhexadecimal. Set the STV as follows:• Input type 2 or 3: 5 V• Input type 4:Input type 4:2v10b104Input type 2 or 3:1v-15ac0Input type 4:210 VAllow the count value on the No. 2 display to fully stabilize, then press theD key to temporarily register the calibration settings.If this count value is outside of the specified range, the No.