Display the data of thismeasurement task in the chart.Only available for scan recipes (see recipeparameter "Mass Mode").Add Total PressureTask48Icon only available for QMG220.Depends on the working area into whichyou drag the tool:•Dragging the tool to the recipe editor:Add a new total pressuremeasurement task to the recipe.•Dragging the tool to the chart:Add a total pressure measurementtask to the recipe. Display the data ofthis measurement task in the chart.QUADERAIconNameDescriptionAdd Working AreaCreate a new working area in the currentrecipe or measurement project.Remove ChartElementDelete a measuring line or a measurementtask from the chart.Remove SectionDelete the selected section.Remove SectionContentDelete all objects of a section, such ascharts and measurement tasks etc.Remove WorkingAreaDelete the current working area.Split SectionBottomSplit the section in half horizontally. A newsection is added to the bottom of thepresent contents.Split Section LeftSplit the section in half vertically.
A newsection is added to the left of the presentcontents.Split Section RightSplit the section in half vertically. A newsection is added to the right of the presentcontents.Split Section TopSplit the section in half horizontally. A newsection is added to the top of the presentcontents.49QUADERA HTML HelpMeasurement Project WindowThe measurement project window is a subarea of the QUADERA® user interface.The measurement window is used to display a measurement project and contains thefollowing areas:50•Recipe selection,•Working area,•Working area selection.QUADERARecipe SelectionThe recipe selection is a subarea of the measurement project window.The recipe selection tab allows you to switch to a recipe within a measurementproject.51QUADERA HTML HelpWorking AreaThe working area is a subarea of the measurement project window or of a recipe.The working area is used for displaying measuring data and for editing the currentlyselected recipe.
The recipe editor is present in every recipe and cannot be removedby the user. Additional display areas serve as containers for the related measuringdata displays. The user can add several display areas to a recipe.52QUADERAWorking Area SelectionThe working area selection is a subarea of the measurement project window.Several working areas are available for each recipe. The working area selection tabsallow you to switch between the displays.53QUADERA HTML HelpDevice Status ViewThe device status view is a subarea of the QUADERA® user interface.The device status view shows the most important device parameters and states.
At aglance, you can see all the important information on the status of the connecteddevice, the measurement and the recipe. The displayed icons and parameters arepredetermined, and cannot be shown and hidden. Use the icons to easily access thebasic functions of the software and the device:IconDescriptionDevice disconnected.Click on the icon to connect the device.Device connected.Click on the icon to disconnect the device.Basic device settings.Click on the icon to view the basic device settings in the"Device Configuration" window, and edit as appropriate.Measurement stopped.Click on the icon to start the measurement.Measurement running.Click on the icon to stop the measurement.Filament off.Click on the icon to switch the filament on.Filament on.Click on the icon to switch the filament off.SEM off.Click on the icon to switch the SEM on.SEM on.Click on the icon to switch the SEM off.Common SEM VoltageCommon operating voltage of the SEM.
Enter the required value or use the spinbuttons. Measurement tasks will use the "Common SEM Voltage" if <<SEMVoltage>> is selected from the dropdown list for the corresponding recipe parameter"SEM Voltage".NOTE:The "Common SEM Voltage" parameter is not available when the FaradayCup has been selected in the basic device settings (Symbol "DeviceConfiguration View").54QUADERAEvent ViewThe event view is a subarea of the QUADERA® user interface.The event view informs you of events that have occurred. Warnings, informations anderrors are displayed in a compact way.Events related to measurement, device and system are listed. The presentationfollows the event view used by the Windows operating system. Refer to the list ofevents for a detailed description.55QUADERA HTML HelpContext MenusContext MenusA context menu is a special menu that opens up when pressing the right mousebutton.
It is context sensitive and always contains commands which are related to thecurrently used screen area.In QUADERA®, you may call up context menus in the following screen areas:Title barContext menu with functions relating to the QUADERA®window.Window titleContext menu with functions relating to the respectivewindow in QUADERA®.Menu barContext menu with functions relating to the menu barsettings.Recipe SelectionContext menu used for adjusting the recipe layout(working area)Frame of thesectionContext menu used for showing and hiding thenavigator.Charts in theworking areaContext menus with functions used for adjusting thechart layout.Context menus in QUADERA®56QUADERACalling Up a Context Menu1.
Position the mouse pointer in the required screen area.2. Press the right mouse button.57QUADERA HTML HelpContext Menu "Title Bar"Click with the right mouse button into the title bar of a window. The context menu"title bar" appears. It contains the following functions:58RestoreRestore the default size for the selected window.MoveMove the window.SizeChange the window size.MinimizeMinimize the window size.MaximizeMaximize the window size.CloseClose the window or the application.QUADERAContext Menu "Menu Bar"Click with the right mouse button into the menu bar.
The context menu "Menu bar"appears. It contains the following functions:•Lock theToolbarsCheckbox: Determine whether or not it ispossible to change the positions of the menubars.•Customize...Open a window which allows you to customizeor create menu bars, and also to configurekeyboard shortcuts.59QUADERA HTML HelpContext Menu "Recipe Selection"Click with the right mouse button into the recipe selection area.
The context menu"Recipe Selection" appears. It contains the following functions:•60Renameselected tabOpen the dialog field "Rename selected tab" torename the tab of the selected working area.QUADERAContext Menu "Frame of the Section"Click with the right mouse button into the frame of the section. The context menu"Frame of the section" appears. It contains the following functions:•ShowNavigatorShow or hide the data navigator.•Renameselected tabOpen the dialog field "Rename selected tab" torename the tab of the selected working area.61QUADERA HTML HelpContext Menu "Chart"Click with the right mouse button into a chart located in the working area.
Variouscontext menus are available, depending on the selected mouse pointer position:•Click on the measuring data display. Context menu "Chart" appears. Itcontains the following functions:• Zoom Out• Reset Zoom• Hold• Legend• Marker (only in MID/MCD recipes)• Display (only in MID/MCD recipes)•Click on the x-axis.
Context menu "Axis X" appears.•Click on the y-axis. Context menu "Axis Y" appears.•Only when the legend is shown: Click on the chart legend. Context menu"Legend" appears.•Only for trending charts in MID/MCD recipes: Click on the right y-axis. Contextmenu "Axis Y2" appears.Context menu "Chart"Click on the measuring data display to open this context menu.•Zoom OutUndo the last zoom action.•Reset ZoomUndo all zoom actions, i.e.
reset the display tothe initial area.•Hold•Enabled (with checkmark):Freeze the display during a measurement(no scaling and scrolling). Measuring dataare still recorded and stored. "Hold On" isautomatically selected when zooming ormoving the cursor.•Disabled (no checkmark):Disable the hold function and resume todisplay the recording of data.
"Hold Off" isusually applied after zooming or moving thecursor.NOTE: There is one hold function for eachworking area.•62Legend•Enabled (with checkmark):The chart legend is displayed.•Disabled (no checkmark):The chart legend is not displayed.QUADERA••MarkerDisplayOnly available for trending charts in MID/MCDrecipes:•Enabled (with checkmark):Displays the individual measuring points andthe connecting lines.•Disabled (no checkmark):Displays the connecting lines of themeasuring points.Only available for MID/MCD recipes.Switch the representation of the chart:•Trend:Continuously display the measuring data asa function of the time (in the selectedsection, see data navigator).•Bar:Display the measuring data in a bar graph atthe time specified by the cursor, see datanavigator.•Table:Display the measuring data in a table at thetime specified by the cursor, see datanavigator.Context menu "Axis X"Click on the x-axis to open this context menu.
It is used to switch thepresentation of the x-axis.•FixedRange of the axis is fixedOnly in scan recipes•AutomaticRange of the axis is adjusted to the measuredvaluesOnly in scan recipes••TimeAbsoluteDisplay the absolute time (date, time)Time RelativeDisplay the relative time (measuring time)Only in MID/MCD recipesOnly in MID/MCD recipesContext menu "Axis Y"Click on the y-axis to open this context menu. It is used to switch thepresentation of the y-axis.•LinearLinear axis scaling•LogarithmicLogarithmic axis scaling•FixedRange of the axis is fixed•AutomaticRange of the axis is adjusted to the measuredvalues63QUADERA HTML HelpContext menu "Axis Y2"This context menu is available only in MID/MCD recipes, and at leastone measurement task is assigned to the second y-axis (via contextmenu "Legend"). Click on the right y-axis to open this context menu.