The "Add-in Manager" windowopens up.4. In the displayed "Available Add-ins" list, click the name of the add-in to beedited, or browse for the required add-in.5. Click [Edit]. The source code is being opened in the VSTA editor for editing.QUADERA® stores each add-in in a main scripting directory. When required the addins are available for all measurement project templates.NOTE:For further information on scripting and the programming environment,please refer to the help file for Microsoft Visual Studio Tools for Applications.Use the Help menu of the VSTA editor.178QUADERACalling Up the VSTA Editor1. Open an existing measurement project, or create a new measurement project.2.
In the Tools menu, point to Programming.3. In the submenu, select Add-in Manager…. The "Add-in Manager" windowopens up.4. In the displayed "Available Add-ins" list, click the name of the add-in to beedited, or browse for the required add-in.5. Click [Edit]. The source code is being opened in the VSTA editor for editing.QUADERA® stores each add-in in a main scripting directory. When required the addins are available for all measurement project templates.NOTE:For further information on scripting and the programming environment,please refer to the help file for Microsoft Visual Studio Tools for Applications.Use the Help menu of the VSTA editor.179QUADERA HTML HelpCalling Up the VSTA Help1. Call up the VSTA editor.2.
In the Help menu, select the required item.or2. Press the F1 key.NOTE:The VSTA Help includes the help for the specific QUADERA® scriptingfunctionality.180QUADERAOperating the Input/Output Module•Quadrupole mass spectrometer QMG220: input/output module IO220•Quadrupole mass spectrometer QMG700: input/output module IO700Use the relevant measurement project as a user interface for testing and servicingthe input/output module (if existing). For this, connect the device with QUADERA®.•Digital Inputs: display the states of the digital input (DI) channels:• Number: identification of the DI channel,• Value: value of the DI channel (On/Off),•Digital Outputs: switch on and off the digital output channels (DO) manually:• Number: identification of the DO channel,• Value: value of the DO channel (On/Off),•Analog Inputs: display the voltage at the analog input (AI):• Number: identification of the AI channel,• Value: numerical value of the voltage at the AI channel.Add a measurement task "Add Analog Input Task" to Scan and MIDmeasurement projects to measure the voltage at a analog input.•Analog Outputs: Change the voltage at the analog output (AO):• Number: identification of the analog output,• Value: numerical value of the voltage at the analog output.181QUADERA HTML HelpWarning and Error MessagesList of EventsThe event view informs you on the occurred events:•"Date/Time" displays the time when the event occurred,•"Source" displays the source that caused the event, e.g.
measurement ordevice,•"Type" displays the event type ("Information", "Warning", "Error"),•"Description" displays a brief message text on the event, and specifieswarnings and errors. This three-part specification contains:• the device specification,• the code letter of the event type ("E" means error, "W" means warning),• the three-digit identification number of the event.List of events sorted by identification numbers:•Warning messages ("W") for the QMG220:• W001: "emission current out of tolerance"• W002: "emission current wide out of tolerance"• W003: "RF-stage temperature is high"•Error messages ("E") for the QMG220:• E001: "external protection triggered"• E002: "emission error"• E003: "filament 1 defect"• E004: "filament 2 defect"• E005: "RF-stage switched off because of excessive temperature"• E009: "ionsource error"• E039: "SEM error"182QUADERAWarning Messages for the QMG220Warning Message QMG220.W001Message text:"emission current out of tolerance"Description:The emission current deviates from the set value.Message occurs when?•The emission current deviates 10% and more than 40 µA from the specifiedvalue.Reasons•The emission current control is faulty,•The pressure in the recipient is unstable.Consequences•Measuring data are corrupt.Solutions•Replace the IS220 when the message occurs repeatedly with stable pressureconditions.183QUADERA HTML HelpWarning Message QMG220.W002Message text:"emission current wide out of tolerance"Description:The emission current essentially deviates from the set value.Message occurs when?•The emission current deviates more than 20% and more than 40 µA from thespecified value.Reasons•The emission current control is faulty,•The pressure in the recipient is unstable.Consequences•Measuring data are corrupt.Solutions•184Replace the IS220 when the message occurs repeatedly with stable pressureconditions.QUADERAWarning Message QMG220.W003Message text:"RF-stage temperature is high"Description:The temperature of the RF stage has exceeded 65°C.Message occurs when?•The temperature of the RF stage exceeds 65°C.Reasons•Cooling of the RF stage is not sufficient because the fan filter pad is blocked,•The fan is defective.Consequences•Measuring data are corrupt.Solutions•Check the fan filter pad,•Repair the fan.185QUADERA HTML HelpError Messages for the QMG220Error Message QMG220.E001Message text:"external protection triggered"Description:The protection for SEM and filament has triggered.Message occurs when?•The specified threshold is exceeded when monitoring the total pressure,•The external protection is triggered,•The total pressure gauge is not properly connected, or the protection input isnot properly connected.Reasons•Connected total pressure gauge: the pressure has exceeded the specifiedthreshold,•The total pressure gauge reports an error,•The total pressure gauge is defective or disconnected,•The total pressure gauge is switched off.Consequences•Filament and SEM are switched on and off depending on the selectedprotection mode.Solutions•Set the protection mode to internal protection.•Connect a total pressure gauge.•Modify the threshold of the total pressure gauge.186QUADERAError Message QMG220.E002Message text:"emission error"Description:An emission error occurred.Message occurs when?•The specified emission current is not reached,•The protection current for the filament is triggered.Reasons•The filament emission current is too low.Consequences•Ionization is impossible,•Measuring data are not meaningful.Solutions•Check the specified protection current for the filament,•Replace the filament.
Refer to the device specific instructions of the QMG (OEMmanual) for further information,•Switch over to the other filament (when existing).187QUADERA HTML HelpError Message QMG220.E003Message text:"filament 1 defect"Description:The filament 1 is broken.Message occurs when?•A broken filament is detected when the device is switched on,•A broken filament is detected when the filament is switched on,•The filament is missing.Reasons•The filament is defective.Consequences•Ionization is impossible,•The filament can not be switched on,•Measuring data are not meaningful.Solutions•Replace the filament. Refer to the device specific instructions of the QMG (OEMmanual) for further information.•Switch over to the filament 2 (when existing).188QUADERAError Message QMG220.E004Message text:"filament 2 defect"Description:The filament 2 is broken.Message occurs when?•A broken filament is detected when the device is switched on,•A broken filament is detected when the filament is switched on,•The filament is missing.Reasons•The filament is defective.Consequences•Ionization is impossible,•The filament can not be switched on,•Measuring data are not meaningful.Solutions•Replace the filament.
Refer to the device specific instructions of the QMG (OEMmanual) for further information.•Switch over to the filament 2 (when existing).189QUADERA HTML HelpError Message QMG220.E005Message text:"RF-stage switched off because of excessive temperature"Description:The RF stage was switched off because the temperature has exceeded 80°C.Message occurs when?•The temperature of the RF stage exceeds 80°C.Reasons•Cooling of the RF stage is not sufficient because the fan filter pad is blocked,•The fan is defective.Consequences•The power supply of the RF stage is switched off,•Measuring data are not meaningful.Solutions•Check the fan filter pad,•Repair the fan,•Switch the device off, and wait until the RF stage has cooled down.
Thenswitch the device on again.190QUADERAError Message QMG220.E009Message text:"ionsource error"Description:An error occurred in the power supply of the ion source.Message occurs when?•The current related to the ion reference voltage is too high,•The current limit is exceeded in degas mode,•The power supply of the IS220 is not ok.Reasons•The feeder of the IS220 is not ok,•The connected ion source is short-circuited,•The specified protection current limit for degas mode is too low.Consequences•The ion source is operating faulty,•Measuring data are not meaningful.Solutions•Check and correct contacts and cable of the IS220 power supply,•Dismount the ion source.