Quadera Pfeiffer программа управления (1248465), страница 18
Текст из файла (страница 18)
Thus, even verysmall peaks can be detected in thebackground signal.•STAIR-T (only peak tasks):Record a mass spectrum withinteger mass jumps.•PEAK-L (only peak tasks):Scan-analog measurement withthreshold value. Only that peaksare recorded and displayed, whichexceed the threshold value.•PEAK-F (only peak tasks):Ion currents are additionallyfiltered using a FIR algorithm.•DetectorTypeDropdown listUsed detector type.•AI ChannelDropdown listNumber of the analog input formeasuring data (only relevant fordetector type AI in analog inputmeasurement tasks).•GainDropdown listParameter valid only for QMG700Measuring range for the analog inputof the QC (only relevant for detectortype EXT_IN_# in analog inputmeasurement tasks).•SEMVoltageNumber > 0Operating voltage for the SEM.Enter the required value or use thespin buttons.The value "0" means that thecommon SEM voltage is applied.
It isdisplayed and set in the device statusview.152QUADERAParameterRangeDescription•Number > 0Parameter valid only for QMG700CP-LevelResponse threshold of the ion counterpreamplifier.•AutorangeModeDropdown listIn QUADERA®, the amplifiermeasuring range can be fixed oradjusted automatically.•FIX:You can select the measuringrange under Amplifier Range.•AUTO:The measuring range is selectedautomatically from all possibleranges.•AUTO-DOWN:The measuring range is selectedautomatically from all rangeslocated between the largest (leastsensitive) range and the RangeLimit.•DetectorRangeDropdown listMeasuring range for the ion current(only relevant for AutorangeMode=FIX)•Range LimitDropdown listMost sensitive measuring range forthe ion current (only relevant forAutorange Mode=AUTO-DOWN)•AnalogSignal FilterDropdown listParameter valid only for QMG700ResolutionInteger number>0•Time constant of analog filter.Mass resolution (small number =clear separation of peaks).
Typicalvalues are approx. 25 for QMG700,and approx. 50 for QMG220.Enter the required value or use thespin buttons.•PauseCalibrateNumber > 0A factor used to multiply the pausetime determined in the device. Inorder to wait for transient effects, nomeasurements are performed duringthe pause time between twomeasurement tasks.Enter the required value or use thespin buttons.•AO ChannelDropdown listNumber of the analog output.Measured values are output to thisAO.153QUADERA HTML HelpParameterRangeDescription•Dropdown listParameter valid only for QMG220AO ModeScales the measurement signal output(0…10 V). Parameter valid only forspecified AO Channel.•AO RangeDropdown listParameter valid only for QMG220Adjusts the measurement range tothe maximum measured signaloutput.
Parameter valid only forspecified AO Channel.•Trip TypeDropdown listParameter valid only for analog inputand total pressure measurementtasks.The trip function can monitormeasurement values of analog inputmeasurements. Each measurementtask provides two trip functions, Aand B, and the related thresholdvalues, Level A and Level B. The tripfunction results are sent to the outputBits DO# A and DO# B.•OFF:The trip function is switched off.•ABS:Use the trip function "ABS" forvacuum monitoring, differentialpressure monitoring, and ascomparator.
Read furtherinformation.•HYST:Use the trip function "HYST" toprevent permanent switching forfluttering signals. Read furtherinformation.•Level ANumberThreshold value A (available only fortrip types ABS and HYST).•DO# AInteger number>0Digital output Bit. The result of thetrip function is output to this DO(available only for trip types ABS andHYST).•Level BNumberThreshold value B (available only fortrip types ABS and HYST).•DO# BInteger number>0Digital output Bit.
The result of thetrip function is output to this DO(available only for trip type ABS).154QUADERAParameterRangeDescription•Dropdown listStatus of the measurement task.State•ENABLE:The measurement task is beingmeasured and the parameters arebeing saved.•SKIP:The measurement task is notbeing measured, but theparameters are preserved.155QUADERA HTML HelpRecipe Parameters: MIDNOTE:New or modified parameters for the measurement tasks are accepted by thedevice only after the measurement has been restarted.The following recipe parameters are available:Task NameMassDwellDetector TypeAI ChannelGain (only QMG700)SEM VoltageCP-Level (only QMG700)Autorange ModeDetector RangeRange LimitAnalog Signal Filter (only QMG700)ResolutionPause CalibrateAO ChannelAO Mode (only QMG220)AO Range (only QMG220)Trip TypeLevel ADO# ALevel BDO# BState156QUADERAParameterRangeDescription•Task NameLetters anddigitsName of the measurement task•MassNumber > 0Mass number for the measurementtask.Enter the required value or use thespin buttons.•DwellDropdown listDwell time for the measurement atthe current mass.In analog input measurement tasks:Dwell time for the measurement ofthe analog input of the QC (onlyrelevant for detector type EXT_IN_#).•DetectorTypeDropdown listUsed detector type•AI ChannelDropdown listNumber of the analog input formeasuring data (only relevant fordetector type AI in analog inputmeasurement tasks).•GainDropdown listParameter valid only for QMG700Measuring range for the analog inputof the quadrupole controller QC (onlyrelevant for detector type EXT_IN_#in analog input measurement tasks).•SEMVoltageNumber > 0Operating voltage for the SEM.Enter the required value or use thespin buttons.The value "0" means that thecommon SEM voltage is applied.
It isdisplayed and set in the device statusview.•CP-LevelNumber > 0Parameter valid only for QMG700Response threshold of the ion counterpreamplifier.157QUADERA HTML HelpParameterRangeDescription•Dropdown listIn QUADERA®, the amplifiermeasuring range can be fixed oradjusted automatically.AutorangeMode•FIX:You can select the measuringrange under Amplifier Range.•AUTO:The measuring range is selectedautomatically from all possibleranges.•AUTO-DOWN:The measuring range is selectedautomatically from all rangeslocated between the largest (leastsensitive) range and the RangeLimit.•DetectorRangeDropdown listMeasuring range for the ion current(only relevant for AutorangeMode=FIX)•Range LimitDropdown listMost sensitive measuring range forthe ion current (only relevant forAutorange Mode=AUTO-DOWN)•AnalogSignal FilterDropdown listParameter valid only for QMG700ResolutionInteger number>0•Time constant of analog filter.Mass resolution (small number =clear separation of peaks).
Typicalvalues are approx. 25 for QMG700,and approx. 50 for QMG220.Enter the required value or use thespin buttons.•PauseCalibrateNumber > 0A factor used to multiply the pausetime determined in the device. Inorder to wait for transient effects, nomeasurements are performed duringthe pause time between twomeasurement tasks.Enter the required value or use thespin buttons.•AO ChannelDropdown listNumber of the analog output.Measured values are output to thisAO.•AO ModeDropdown listParameter valid only for QMG220Scales the measurement signal output(0…10 V). Parameter valid only forspecified AO Channel.158QUADERAParameterRangeDescription•Dropdown listParameter valid only for QMG220AO RangeAdjusts the measurement range tothe maximum measured signaloutput.
Parameter valid only forspecified AO Channel.•Trip TypeDropdown listThe trip function can monitormeasurement values of MID- andMCD measurements, total pressuremeasurements, and analog inputmeasurements (AI). Eachmeasurement task provides two tripfunctions, A and B, and the relatedthreshold values, Level A and Level B.The trip function results are sent tothe output Bits DO# A and DO# B.•OFF:The trip function is switched off.•ABS:Use the trip function "ABS" forvacuum monitoring, differentialpressure monitoring, and ascomparator.
Read furtherinformation.•HYST:Use the trip function "HYST" toprevent permanent switching forfluttering signals. Read furtherinformation.•Level ANumberThreshold value A (available only fortrip types ABS and HYST).•DO# AInteger number>0Digital output Bit. The result of thetrip function is output to this DO(available only for trip types ABS andHYST).•Level BNumberThreshold value B (available only fortrip types ABS and HYST).•DO# BInteger number>0Digital output Bit. The result of thetrip function is output to this DO(available only for trip type ABS).•StateDropdown listStatus of the measurement task.•ENABLE:The measurement task is beingmeasured and the parameters arebeing saved.•SKIP:The measurement task is notbeing measured, but theparameters are preserved.159QUADERA HTML Help160QUADERARecipe Parameters: MCDNOTE:New or modified parameters for the measurement tasks are accepted by thedevice only after the measurement has been restarted.An MCD measurement project is based on an MID measurement project and includesadditional evaluation to yield the components' concentrations of a gas mixture(quantitative gas analysis).
Thus, an MCD measurement project always contains ameasurement recipe and an analysis matrix which contains the gas specific calibrationfactors. The MCD measurement project is closely related to a gas specific calibrationwhich is needed to determine the relevant calibration factors. Note that you shouldperform the gas specific calibration directly before the MCD measurement project.Analysis matrix (MCD Matrix Editor)The analysis matrix connects the MID measurement recipe with the gas components:•The rows of the analysis matrix are defined by the name of the gascomponents for the quantitative analysis.•Use the "Add MCD Task" tool to add a row (gas component) to the matrix.NOTE:In the MCD project, use identical names for the gas components as used inthe gas specific calibration.
Otherwise, QUADERA® cannot refer to thecomponents. As an example, QUADERA® does not equate "O2" and"Oxygen".•The columns of the analysis matrix are defined by the measurement recipe,and correspond to the mass numbers in the measurement tasks.•The individual matrix elements represent the gas specific calibration factors forthe quantitative gas analysis. The calibration factors are determined in the gasspecific calibration and stored in the MCD Calibration Factor Library.161QUADERA HTML HelpRecipe Parameters: Total PressureThis chapter is valid for the QMG220 only.NOTE:New or modified parameters for the measurement tasks are accepted by thedevice only after the measurement has been restarted.The following recipe parameters are available::Task NameDwellDetector TypeTrip TypeLevel ADO# ALevel BDO# BStateParameterRangeDescription•Task NameLetters anddigitsName of the measurement task.•DwellDropdown listDwell time for the measurement ofthe total pressure.•DetectorType(not editable)Detector type "TotalPressure":installed total pressure gauge162QUADERAParameterRangeDescription•Dropdown listThe trip function can monitormeasurement values of total pressuremeasurements.
Each measurementtask provides two trip functions, Aand B, and the related thresholdvalues, Level A and Level B. The tripfunction results are sent to the outputBits DO# A and DO# B.Trip Type•OFF:The trip function is switched off.•ABS:Use the trip function "ABS" forvacuum monitoring, differentialpressure monitoring, and ascomparator. Read furtherinformation.•HYST:Use the trip function "HYST" toprevent permanent switching forfluttering signals.